HS-420RT Accelerometer 4-20mA velocity with temperature output- M12

Vibration Monitoring with Hansford Sensors

Our advanced vibration transmitters, or loop powered sensors, are engineered for use with the latest PLC (programmable logic controller), BMS (building management system), or SCADA (supervisory control and data acquisition) technologies, offering sophisticated solutions for condition monitoring and preventative maintenance.

The family is available with a variety of different cables.

 

Key Features:

  • 4-20mA output proportional to velocity
  • Temperature (Pt100)
  • Various mountings
  • Sealed to IP67
  • Low cost cable assembly


Typical Applications:
Fans, pumps, motors, gearboxes, compressore, centrifuges, conveyors, Air handlers, rolls, dryers, presses, cooling, VAC, spindles, machine tooling, process equipment

 

Technical Performance
Mounted Base Resonance: 5kHz min
Velocity Ranges: see: ‘How To Order’ table ± 10% Nominal 80Hz at 22°C
Frequency Response: 10Hz (600 kcpm) to 1kHz (60kcpm) ± 5% – ISO10816
Isolation: Base isolated
Range: 50g peak
Temperature Output: Pt100 (100Ohm)
Transverse Sensitivity: Less than 5%

 

Mechanical
Case Material: Stainless Steel
Sensing Element/Construction: PZT/Compression
Mounting Torque: 8Nm
Weight: 150gms (nominal)
Screened Cable Assembly: HS-AA005 – non-booted
HS-AA068 or HS-0069 – booted
Mounting Threads: see: ‘How To Order’ table

 

Electrical
Current Output: 4-20mA DC proportional to Velocity Range
Supply Voltage: 15-30 Volts DC (for 4-20mA)
Settling Time: 2 seconds
Output Impedance: Loop Resistance 600 Ohms max. at 24 Volts
Case Isolation: >108 Ohms at 500 Volts

 

Environmental
Operating Temperature Range: -25 to 90ºC
Sealing: IP67
Maximum Shock: 5000g
EMC: EN61326-1:2013

 

Typical Frequency Response (at 100mV/g):

 