In
Portable Vibration Calibrator AT-2030
☰
filters
AT2030 Portable Calibration Shaker
A portable calibration shaker that produces a known, traceable vibration signal in g’s, mils or ips from a battery-powered case you can carry. Calibrate and check accelerometers, proximity probes, monitoring systems and wiring on site, from 5 Hz to 10 kHz, without the complexity of advanced features.
How it works
A known input, held by closed-loop control.
A quartz reference accelerometer under the shaker head measures the real motion. It’s calibrated back-to-back against a laser-calibrated primary reference.
Digital closed-loop control compares that measurement with your target and adjusts the drive, while PureWave™ distortion correction actively cancels up to 30 harmonics, so the level stays put as you change frequency or mount a heavier sensor. Mount your sensor, read it on your own meter, analyzer or monitoring system, and compare it with the level on screen. THD and mass load are shown live, so a loose mount or an overloaded head shows up straight away.Specifications
The details.
| Performance | |
|---|---|
| Frequency range | 5 Hz – 10,000 Hz (300 – 600,000 RPM), 100 g payload |
| Maximum payload | 800 g (limited to 0.1 in / 2.54 mm peak displacement at low frequency) |
| Waveform | Sine |
| Maximum output | 20 g pk (196 m/s²) · 20 in/s pk (508 mm/s) · 0.20 in pk-pk (5.1 mm), bare table |
| PureWave™ distortion correction | Active harmonic cancellation, corrects up to 30 harmonics |
| Accuracy | |
| Acceleration | ±3.2 % (5 – 9 Hz) · ±2.2 % (10 Hz – < 5 kHz) · ±3.0 % (5 – 10 kHz) |
| Displacement | ±2.2 % (30 – 150 Hz) |
| Amplitude linearity | < 1 % up to 10 g pk (100 g payload, 100 Hz) |
| Waveform distortion | < 1 % THD typical up to 5 g pk (100 g payload, 30 Hz – 2 kHz) |
| Reference | Precision quartz reference accelerometer, laser-calibrated primary reference |
| Readout | |
| Acceleration | g pk, m/s² pk, g RMS, m/s² RMS |
| Velocity | in/s pk, mm/s pk, in/s RMS, mm/s RMS |
| Displacement | mils p-p, mm p-p, µm p-p |
| Frequency | Hz, RPM |
| System | |
| Architecture | Embedded Linux, FPGA-based signal generation and acquisition, 24-bit codec |
| Signal processing | DFT amplitude and phase, real-time harmonic analysis, automatic gain ranging |
| Memory | 16 GB internal, microSD slot |
| Safety | Amplitude limiting by frequency, 200 mils p-p over-displacement protection, automatic shutdown on alarm |
| Network remote control | Ethernet or Wi-Fi · plain-text script commands over TCP · VNC screen control (see how) |
| Power | |
| Battery | 12 V DC, 6 Ah LiFePO4 · 5,000 cycles at 80 % depth of discharge, or 10 years |
| Charging | 1 hour · 100–240 V AC, 50–60 Hz integrated charger, IEC C13 cord |
| Battery life | Up to 18 h low output · ~4–12 h typical testing · ~1 h continuous high output |
| Physical | |
| Display | 4.3″ capacitive TFT LCD, 480 × 272, high-brightness, anti-glare |
| Controls | 2 dials and touchscreen |
| Sensor mounting | ¼-28 thread |
| Dimensions (H × W × D) | 10.6 × 9.7 × 6.9 in (27 × 24.6 × 17.4 cm) |
| Weight | 13.0 lb (6.24 kg) |
| Operating temperature | 32 – 122 °F (0 – 50 °C) |
| Certifications, accessories & support | |
| Certifications | A2LA accredited · ISO/IEC 17025:2017 · NIST traceable · ANSI/NCSL Z540 traceable · EMC EN 61326-1 · LVD EN 61010-1 · RoHS |
| Included | Power cable, ¼-28 stud, 10-32 UNF stud, 2-56 UNC, 6-32 UNC and 10-32 UNF adapters, universal velocity and accelerometer adapter discs, short-handle wrench, USB drive with CSV import templates and manual |
| Optional | Proximity probe adapter kit (digital or manual micrometer), Chadwick-Helmuth velocimeter cable, triaxial accelerometer adapter |
| Warranty & support | 2 years, including drift and accuracy · training webinars and email support |